A Bowie-Dick test is used in pre-vacuum type (or dynamic air removal) sterilizers. They are used to detect air leaks and inadequate air
removal and consist of folded 100% cotton surgical towels that are clean and preconditioned. A commercially available Bowie-Dick-type
test sheet should be placed in the center of the pack.
The test pack should be placed horizontally in the front, bottom section of the sterilizer rack, near the door and over the drain, in an
otherwise empty chamber and run at 134°C for 3.5 minutes. The test is used each day the vacuum-type steam sterilizer is used, before
the first processed load.
Air that is not removed from the chamber will interfere with steam contact. Smaller, commercially available disposable test packs (or
process challenge devices) have been devised to replace the stack of folded surgical towels for testing the efficacy of the vacuum
system in a pre vacuum sterilizer.
They should be representative of the load and simulate the greatest challenge to the load. Sterilizer vacuum performance is acceptable
if the sheet inside the test pack shows a uniform color change. Entrapped air will cause a spot to appear on the test sheet, due to the
inability of the steam to reach the chemical indicator. If the sterilizer fails the Bowie-Dick test, do not use the sterilizer until it is
inspected by the sterilizer maintenance personnel and passes the Bowie-Dick test.
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
R8,50
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
Be the first to review “Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)” Cancel reply
Related products
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R20,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R395,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R24 840,00

Rubber Tip for All walking Frames
Tray instruments with Cover 305 x 205 x 50mm
Blood Pressure Meter Aneroid on mobile Stand (Telestand)
Mattress Hospital Anti Bedsores 2-Section medium density
ECG Portable PM10 with Bluetooth data transmission, LI Battery
Walking Stick Quad Straight
Blood Pressure Aneroid single-handed with shatterproof gauge
ECG Electrodes Pack of 50’s Neonate
Scissors budget Rescue 19cm, 1’s (Moq)
Trolley Emergency 5 drawers ABS body
AED CU-SP1 _ Pads Pediatric only
Suction unit Handheld 300ml Bottle 39Kpa
Simulator IV Training arm
Cold Sterilizer Steritech 5L
Splint wooden 1’s- Moq
Bath Seat Sunken
Bath Aid Bench
Autoclave BKM-K18N Table Top
Blood Pressure Meter Aneroid Wall Mount Square swivel
Diathermy _ Handle and Finger-switch Re-use 








Reviews
There are no reviews yet.