A Bowie-Dick test is used in pre-vacuum type (or dynamic air removal) sterilizers. They are used to detect air leaks and inadequate air
removal and consist of folded 100% cotton surgical towels that are clean and preconditioned. A commercially available Bowie-Dick-type
test sheet should be placed in the center of the pack.
The test pack should be placed horizontally in the front, bottom section of the sterilizer rack, near the door and over the drain, in an
otherwise empty chamber and run at 134°C for 3.5 minutes. The test is used each day the vacuum-type steam sterilizer is used, before
the first processed load.
Air that is not removed from the chamber will interfere with steam contact. Smaller, commercially available disposable test packs (or
process challenge devices) have been devised to replace the stack of folded surgical towels for testing the efficacy of the vacuum
system in a pre vacuum sterilizer.
They should be representative of the load and simulate the greatest challenge to the load. Sterilizer vacuum performance is acceptable
if the sheet inside the test pack shows a uniform color change. Entrapped air will cause a spot to appear on the test sheet, due to the
inability of the steam to reach the chemical indicator. If the sterilizer fails the Bowie-Dick test, do not use the sterilizer until it is
inspected by the sterilizer maintenance personnel and passes the Bowie-Dick test.
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
R8,50
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
Be the first to review “Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)” Cancel reply
Related products
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R31 895,07
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
Autoclave 16L Table top with Micro-processor Control (Class2)
R48 983,65
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R440,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R11 955,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more

Blood Pressure Meter Wall Mount or Desk
Spirometer SP10
Antiseptic Solution 50ML or 100ML Hi-Care, 1’s (Moq)
Defibrillator Life7, Paramedic series, Pacemaker & NiBp/Spo2, 5-Lead Cable, Printer (Free case valued R3,200))
Sterilizer Ultraviolet Violet 15W _ Bulb only
Shopper Quad Aluminum Front Basket
Overbed Table Flat wooden Top, height Adjust, tilting or non-tilting
AED Saver1, Defibrillator Fully automatic Complete with Carry Case, Public access
ECG 12-Lead model Edan SE-1200 _ Cable only
Capnograph Portable model CA10M – Intubation Tube only
Commode and Shower Combo, 155Kg capacity
Blood Pressure Aneroid single handed
Blood Fluid 1-Tube warmer BFW1000
Walking frame Foldable Front Wheels size Large (ROLLATOR Foldable)
Doppler Fetal BF500+ _ Adaptor only 








Reviews
There are no reviews yet.