A Bowie-Dick test is used in pre-vacuum type (or dynamic air removal) sterilizers. They are used to detect air leaks and inadequate air
removal and consist of folded 100% cotton surgical towels that are clean and preconditioned. A commercially available Bowie-Dick-type
test sheet should be placed in the center of the pack.
The test pack should be placed horizontally in the front, bottom section of the sterilizer rack, near the door and over the drain, in an
otherwise empty chamber and run at 134°C for 3.5 minutes. The test is used each day the vacuum-type steam sterilizer is used, before
the first processed load.
Air that is not removed from the chamber will interfere with steam contact. Smaller, commercially available disposable test packs (or
process challenge devices) have been devised to replace the stack of folded surgical towels for testing the efficacy of the vacuum
system in a pre vacuum sterilizer.
They should be representative of the load and simulate the greatest challenge to the load. Sterilizer vacuum performance is acceptable
if the sheet inside the test pack shows a uniform color change. Entrapped air will cause a spot to appear on the test sheet, due to the
inability of the steam to reach the chemical indicator. If the sterilizer fails the Bowie-Dick test, do not use the sterilizer until it is
inspected by the sterilizer maintenance personnel and passes the Bowie-Dick test.
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
R8,50
Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)
Be the first to review “Autoclave Test Sheet _Bowie Dick Test Sheet 1’s (Moq)” Cancel reply
Related products
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R440,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R20,00
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
R4 987,95
Sterilization/ Autoclaves, UV Sterilizers, Disinfectants & more
Autoclave 16L Table top with Micro-processor Control (Class2)
R48 983,65

Defibrillator Life7 _ 10-Lead ECG Cable only
Capnograph Portable model CA10M
Dressing Trolley Epoxy coated 2 Shelves
EEG Brain mapping machine
Capnograph Portable model CA10M – Intubation Tube only
Delivery Bed mobile Hi-Low, Trendelenburg Hydraulic(Obstetric)
Bandage Adhesive Elastic-Heavy Sport 25-75mm x4.5m, 1’s Hi Care ( Moq)
Toilet or Shower Grab Bar 1’s U-shaped
Fetal Doppler BF600+, 2Mhz portable with up 4hrs Battery operation
Trolley Anesthesia 1 drawer Stainless Steel
Cottonwool Non-Interleaved - 500g Hi-Care (Moq)
AED CU-SP1 _ Pads adult only
CPR Mouthpiece 1’s (Moq)--Hi Care
ECG Machine 12-Lead 8”inch Screen with Interpretation _ Paper only (Copy)
Patient Recovery Stretcher, Reverse Trendelenburg, Oxygen Cylinder Holder
Safety Pins bundle of 10’s (Moq)
Wheelchair Manual with detachable Armrest/Footrest (Steel-PVC)
Obstetric Bed(Delivery) Manual Adjustments via double rachet
Crutch Elbow Aluminum With rubber Tip, 1’s , Size Medium (1.26m extended)
Bath Aid Bench
Surgical Tray Dressing Large 








Reviews
There are no reviews yet.